11. Technical data
Classification acc. to Directive 93/42/EEC
Art. no.
Designation
GA676
Battery
GA672
Acculan 3Ti drill and reamer
GA672
DC voltage
Rated current
Max. power
Rotational direction
Speed
Weight (operational)
Dimensions (L x W x H)
Applied part
EMC
Conforming to standard
The product was subject to 350 preparation cycles as a
test by the manufacturer and has passed.
GA676
Cell type
DC voltage
Nominal capacity
Weight
Dimensions (L x W x H)
EMC
Conforming to standard
22
Class
I
IIa
Acculan 3Ti drill and
reamer
9.6 V
approx. 3 A
250 W
Clockwise/counterclock-
wise (right/left) rotation
-1
min. 0 min
to max.
-1
26 000 min
approx. 1.25 kg
approx. 107 mm x
47 mm x 200 mm
Type BF
IEC/DIN EN 60601-1-2
IEC/DIN EN 60601-1
Battery
NiMH
9.6 V
1.95 Ah
approx. 0.425 kg
approx. 140 mm x
43 mm x 50 mm
IEC/DIN EN 60601-1-2
IEC/DIN EN 60601-1
11.1 Operating mode
Operating
Operation with non-periodic load
mode
and speed changes (type S9 pursu-
ant to IEC EN 60034-1)
Drilling (clockwise/counterclockwise
rotation):
60 second application,
60 second pause
20 repetitions
30 min cooling time
Reaming (clockwise/counterclock-
wise rotation):
30 second application,
30 second pause
8 repetitions
30 min cooling time
Saw mode with GB660R:
30 second application,
60 second pause
3 repetitions
30 min cooling time
11.2 Ambient conditions
Operation
Temperature
10 °C to 27 °C
Without con-
densation
Relative
30 % to 75 %
humidity
Atmospheric
700 hPa to
pressure
1 060 hPa
Storage and
transport
-10 °C to 40 °C
Maximum rec-
ommended
duration of stor-
age is 3 months,
no direct sun-
light
10 % to 90 %
500 hPa to
1 060 hPa
loading