ENGLISH
RESUSCITATION WITH MASTER FLUX
KIT Art. 1504/A (Not included) (Figure 11)
RESUSCITATION procedures WITH MASTER FLUX must be
carried out with the %N2O knob that regulates the percentage
of nitrous oxide in position 0%.
SLOWLY open the oxygen cylinder valve (turn the valve
counter-clockwise); make sure that the nitrous oxide
cylinder is closed (turn the valve clockwise).
Take the Master Flux mask kit off by sliding it off of the spiral tube.
Insert the relative fitting (30) into the spiral tube.
Remove the bag (F) and put the cap (T).
Connect the end of the oxygen treatment tube (31) to the fitting (30).
Connect the other end of the oxygen treatment tube to the bottom stud of the resuscitation bag (32) through the
relative coupling.
Connect the mask (33), to the valve on the resuscitation bag.
Ruotare la manopola FLOW per erogare Ossigeno al flusso desiderato, CONTROLLARE I FLUSSI
VISUALIZZATI.
Turn the FLOW knob to deliver Oxygen at the required flow rate, CHECK THE FLOWS DISPLAYED ON THE
FLOW COLUMNS.
The nitrous oxide column E=0
The oxygen column D= SET FLOW
Place the mask on the patient.
Act on the resuscitation bag in accordance with the resuscitation instructions.
Once use of the device is finished it is IMPORTANT to close the valve on the Oxygen cylinder (by turning it
clockwise).
FLOW
Figura 11
21