ENGLISH
7.1
Instructions for use
Before using the unit-mounted MASTER FLUX version, lock the wheels that have brakes on
them by pressing the tabs downwards.
If it is necessary to move the equipment (for example from one room to another), disengage
the brakes on the wheels, grip the relative handle and pull the device avoiding impact and
sudden movements that could cause it to fall accidentally.
Only move and place Master flux on flat and smooth surfaces.
Make sure the cylinders are stable and secured correctly.
Close the cylinders before moving the device.
Do not use the handles to lift the device.
When using MASTER FLUX in its wall-mounted version, it is possible to move and position
the flowmeter box as required by gripping the relative handle.
DO NOT LEAN ON AND DO NOT REST OBJECTS ON THE FLOWMETER BOX (BOTH
WALL AND UNIT-MOUNTED VERSIONS).
DO NOT FORCE THE ARM OF THE WALL-MOUNTED VERSION.
SLOWLY open the valves on the O
clockwise (Figura 6 Pos L).
Read the pressure on the pressure reducer pressure gauges, (see TECHNICAL DATA) to
estimate the autonomy of the oxygen cylinder.
Choose the most suitable mask for the patient.
Place the mask on the patient fitting it on well around his/her nose.
Secure the mask to the loop on the back of the head rest of the dental chair.
Make sure that the tubes for the mask circuit do not get bent or chocked.
The operator must be stationed in such a way that he/she can visually control:
-
The rubber bag
-
The flow gauges
During the procedure constantly check the flows displayed on the columns
(Figure 10 Pos D, E).
Before leaving the patient, even briefly, shut off sedation.
Figure 10
Make sure that the %N
Using the FLOW knob (Pos A) administer a flow of 3 Lt/min of oxygen for 30 seconds.
INSTRUCT THE PATIENT TO BREATHE IN AND OUT FROM THE NOSE.
Use the FLOW knob (Pos A) to adjust the flow.
Gradually adjust the flow, check the status of the bag (Pos F) which must take on a shape
that we will define as half full half empty.
Make the patient breathe for at least one minute before evaluating the flow.
18
and N
2
O knob (Pos C) index displays 0.
2
O cylinders turn the valves counter-
2