Door regulation
To regulate the door lock pressure, it is necessary to adjust the regulator of the same
door using the double function wrench (extraction tray + wrench) (FIG.10) provided
(spare code DANA008). Turn anti clockwise (pos.A) to increase lock pressure. Turn
clockwise (pos.B) to decrease the lock pressure
Substitution printer paper
Thermographic paper roll 57 mm wide (spare code DNVA363). Thermographic paper can
be printed on one side only. Open the printer cover by pushing the button place the paper
roll by paying attention to the rotation direction, pull the paper until it comes out and close
the cover; the printer is ready for printing.