Insert plug of electronic motor into
the socket at the control unit.
6.3.5
Preparing the tubing set
Installation of the tubing set REF 6022 according to the following guidelines.
Use only Nouvag tube sets REF 6022 for infiltration, otherwise the correct function cannot be guaranteed.
Check the expiry date of the tubing set and ensure that the packaging is not damaged. Using
non-sterile tubing sets can result in serious infections and worst case can be fatal.
When inserting the tubing set, observe the arrow on the cover of the pump compartment. It
indicates the flow direction of the infiltration solution.
The integrated peristaltic pump is used for the infiltration of the tumescent solution.
Do not regulate the amount of irrigation fluid using the roller clamp on the tubing set; with the LipoSurg,
this is regulated instead using the integrated pump. For this reason, make sure to open the roller clamp as
far as it will go (refer also to 7.2 Control elements).
12
Full assembled Conform Cannula Handpiece, ready to use. (Connect suction
tube with Conform Cannula Handpiece and your vacuum pump system).
Operating instructions LipoSurg, REF 31648, V48/18