Service
Maintenance Instructions
Cleaning
Keep the FCT clean from swarf and dust.
After use, remove any trace of oil, and dust from the FCT.
Use an anti-static cleaning cloth to remove dust from the FCT.
Avoid using harsh detergents to clean the FCT.
Calibration
Calibration is a set of operations that compares the values indicated by a measuring system to the corresponding values
realized by external standards.
The accuracy of electronic drifts with time and temperature, which can result in measurement inaccuracy.
Calibration restores the FCT components to their specified accuracy and ensures that the device still meets Desoutter
standards.
Desoutter recommends performing a complete calibration at least once every year. You can shorten this interval based on
the demands of your application.
To do a complete calibration, please contact a certified Desoutter Service Technician or a Force Calibration Laboratory
(ISO/IEC 17025 Accredited Laboratory recommended).
A
4 holes M5x15 on Ø98
C
Connection to Delta xD
E
Force applied for the calibration
For more information about the calibration process, refer to the certificate delivered with the product.
09/2020
A
C
B
Connect the calibration equipment to these threads:
B
FCT 30 M12x1.25 / FCT60 M16x1.5
Install the FCT with four fixing bolts M5x18 class 8.8:
D
tightening torque is 1 Nm (only for calibration purposes)
9 / 216
Product Instructions
E
D