Light intensity on optical axis
The effective light intensity for bilirubin dissociation E
depends on how far away the lamp housing (lower edge) is.
If a higher intensity of light is required, the white light lamps
may be replaced with blue phototherapy lamps (by
DrägerService or trained service personnel). The hours used
by the two additional phototherapy lamps will not be
registered by the hour meter.
Maximum values
Distance
cm
4 blue light
2
lamps
E
mW/cm
bi
6 blue light
lamps
Distribution of light intensity
This measurement was performed at a vertical distance of
40 cm on a bed surface measuring 50 x 30 cm.
Maximum value = 100 %
Intensity specified as a percentage of the maximum value.
Precision: ±5 %.
63
82
91
82
67
88
100
88
63
82
91
82
10 cm 10 cm 10 cm 10 cm 10 cm
Average values on working surface given:
4 blue light lamps: 1.3 mW/cm
6 blue light lamps: 1.8 mW/cm
Disposal
For countries subject to the EU-Directive 2002/96/EC:
This device is subject to EU Directive 2002/96/EC
(WEEE). It is not registered for use in private
households, and may not be disposed of at municipal
collection points for waste electrical and electronic equipment.
Dräger Medical has authorized a firm to dispose of this device
in the proper manner: for more detailed information, please
contact your local Dräger Medical organization.
28
bi
30
40
2.4 ±0.4
1.6 ±0.3
3.4 ±0.6
2.3 ±0.4
63
10 cm
67
10 cm
63
10 cm
2
2
50
1.2 ±0.2
1.6 ±0.3
Instructions for Use Photo-Therapy 4000