TROUBLESHOOTING
Symptom
The display does not light up when
the unit is switched on
Handpiece without pressure or with
a low pressure when the foot switch
is activated
Pressure not adjustable
Pressure reduction during operation
Handpiece without pulse when the
foot switch is actived
Pressure too high
64
Solutions
•
Check the different electrical connections
•
Check the mains
•
Check/replace the fuses and replace them only with the type indicated on the back of the unit
•
Send the unit to your authorized EMS repair center
•
Check the different air connections
•
Check the pressure available for the unit
•
Check that the manometer is indicating more than 1.5 bar
•
Check the compressor
•
Check the connection between the handpiece and unit
•
Test the unit with another handpiece
•
Send the unit to your authorized EMS repair center
•
Check the different air connections
•
Check the pressure available for the unit
•
Check that the manometer is indicating more than 1.5 bar
•
Check the compressor
•
Send the unit to your authorized EMS repair center
•
Check the different air connections
•
Check the compressed air supply
•
Contact your authorized EMS repair center
•
Check that the "Validate" indicator light is on
•
Check the connection of the foot pedal
•
Check the different connections of the unit
•
Check the different supplies of the unit
•
Test the unit with another handpiece
•
Send the unit to your authorized EMS repair center