GIMA SpA
5.1 GENERAL
Equipment Description:
H.F. Electrosurgery Equipment for monopolar minor surgery. The unit provides the facilities to fine preset the output power level
and choose the coagulating effect through the appropriate Crest Factor.
Cabinet:
Bench-top, metal housing, with the frontal and rear panel covered by lexan sheet.
Height: 100 mm
Equipment Mains Input:
Selectable 115 or 230 V~, 50/60 Hz.
Maximum Current:
0,5 A (230V~) or 1A (115V~)
IEC socket with in-line fuses, voltage selector and mechanical mains switch.
Displays:
LED Digital Displays provide the indication of the preset output level as follows:
CUT:
0 – 50
Output Power Control by two keys
The use of the control keys allows the user to control very accurately the output power since it is possible to vary the
preset output power in step as low as just 1 watt.
Max Output Power:
CUT:
50 W +/- 10%
Reference Load Resistance: 400Ω
No Load Max Voltage:
900 Vpp +/-10%
Crest Factor:
CUT:
1.5
Output Frequency: 600 kHz +/- 10%
Modulation Frequency:
CUT:
NONE
Allowed output time:
Duty Cycle:
10 seconds output-25 seconds pause
Manuale d'Istruzione / Instruction's Manual / Manual de Instrucciones
5. TECHNICAL SPECIFICATIONS
Width:
180 mm
Depth:
BLEND:
0 – 45
BLEND:
45 W +/- 10%
BLEND:
1.8
BLEND:
10 kHz
MA126IGBa
180 mm
Weight: 2,8 Kg
COAG:
0 - 40
COAG:
40 W +/- 10%
COAG:
2.5
COAG:
10 kHz
DIATERMO 106
31