Data Transmission; Technical Specifications - Gima OXY-10 Guía De Uso

Ocultar thumbs Ver también para OXY-10:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 54

3.5 Data transmission

This oximeter has the function of wireless data transmission. The user could
effectively transmit the data to computer through the wireless communication
module. Refer to the (Oximeter Data Manager) for detailed information.

4. Technical Specifications

A. Technique: dual-wavelength LED sensor,
LED sensor wavelength:
Red light: 663 nanometers,
Infrared light: 890 nanometers.
Maximal optical output power:
less than 1.5mW maximum average.
B. SpO
measurement
2
Measuring range: 70%~100%
Measuring accuracy:
not greater than 3% for SpO
Note: Accuracy defined as root-mean-square value of deviation
according to ISO 9919.
SpO
alarm low limit range:
2
85%~99% (default 90%)
C. Pulse Rate measurement
Measuring range: 30bpm~240bpm
Measuring accuracy: ±2bpm or ±2% (whichever is greater)
Pulse Rate alarm range:
high limit: 100~240bpm (default 120bpm)
low limit: 30~60bpm (default 50bpm)
D. Perfusion Index (PI) Display
Range: 0.2%~20%
E. Audible &visual alarm function
When measuring, if SpO
alarm limit, the device will alarm automatically and the value which exceeds
limit on the screen will flash. The oximeter will shut down automatically
in 8 seconds with no signal.
F. Display mode: Color OLED Display
G. Power supply requirement:
2 x LR03 (AAA) alkaline batteries
Supply voltage: 3.0VDC
Operating current: ≤40mA
ENGLISH
range from 70% to 100%
2
value or pulse rate value exceeds the preset
2
28
Tabla de contenido
loading

Tabla de contenido