21
4. TECHNICAL SPECIFICATIONS
4.1 Main performance
A. SpO
value display.
2
B. Pulse rate value display, bar graph display.
C. Pulse waveform display.
D. Low-voltage indication: low-voltage indicator appears before working abnormally which is due to
low-voltage.
E. Screen brightness can be changed.
F. A pulse sound indication.
G. With alarm function.
H. With SpO
value and pulse rate value of storage, the stored data can be uploaded to computers.
2
I. It can be connected with an external oximeter probe.
J. Real-time data can be transmitted to computers.
K. Review function.
L. Clock function.
4.2 Main parameters
A. Measurement of SpO
Measuring range: 0%~100%.
Accuracy: When the SpO
the permission of absolute error is ±2%, below 70% unspecified.
B. Measurement of pulse rate
Measuring range: 30bpm~250bpm Accuracy: ±2 bpm or ±2% (select larger).
C. Resolution
SpO
: 1%, Pulse rate: 1bpm.
2
D. Measurement Performance in Weak Filling Condition:
SpO
and pulse rate can be shown correctly when pulse-filling ratio is 0.4%.
2
SpO
error is ±4%, pulse rate error is ±2 bpm or ±2% (select larger).
2
E. Resistance to surrounding light:
The deviation between the value measured in the condition of man-made light or indoor natural light
and that of darkroom is less than ±1%.
F. Power supply requirement: 2.6 V DC~3.6V DC.
G. Optical Sensor
Red light (wavelength is 660nm, 6.65mW) Infrared (wavelength is 880nm, 6.75mW).
H. Adjustable alarm range:
SpO
: 0%~100%.
2
Pulse Rate: 0bpm~254bpm.
5. INSTALLATION
5.1 View of the front panel
Figure 2. Front View
2
measuring range is 70%~100%,
2
ENGLISH