2.5 Caution
A. The finger should be placed properly (see the attached illustration of this
manual), or else it may cause inaccurate measurement.
B. The SpO
sensor and photoelectric receiving tube should be arranged in a
2
way with the subject's arteriole in a position there between.
C. The SpO
sensor should not be used at a location or limb tied with arterial
2
canal or blood pressure cuff or receiving intravenous injection.
D. Do not fix the SpO
pulsation and inaccurate measure of SpO
E. Make sure the optical path is free from any optical obstacles like rubberized
fabric.
F. Excessive ambient light may affect the measuring result. It includes fluores-
cent lamp, dual ruby light, infrared heater, direct sunlight and etc.
G. Strenuous action of the subject or extreme electrosurgical interference may
also affect the accuracy.
H. Testee can not use enamel or other makeup on the finger.

3. TECHNICAL SPECIFICATIONS

A. Display mode: LCD Display
B. Power supply requirement:
1.5V (AAA size) alkaline battery × 2
Supply voltage: 3VDC
Operating current: <=15mA (backlight off)
C. SpO
Parameter Specifications
2
Measuring range: 35%~99%
Accuracy: ±2% (during 75%~99%)
±3% (during 50%~75%)
SpO
alarm:
2
Lower limit: 90%
D. Pulse Rate Parameter Specifications
Measuring range: 30bpm~240bpm
Accuracy: ±2bpm or ±2% (whichever is greater)
Pulse Rate alarm: Upper limit: 120bpm
Lower limit: 50bpm
E. Blood Perfusion Display
Range: 0~20%
F. Resistance to interference of surrounding light
The difference between the value measured in the condition of indoor natural
light and that of darkroom is less than ±1%.
sensor with adhesive or else it may result in venous
2