- Use the up (E) and down (G) buttons to set the preferred parameter values.
6. Press button (H) to start the unit.
The current value of the speed and the time remaining to the end of centrifuging
will appear on the display.
7. To stop the machine before the set time has elapsed press button (H).
8. To open the unit cover press the (C) button. Caution: Opening the cover is
possible only after the unit has come to a complete stop.
9. Remove the tubes with the centrifuged samples.
10. Turn off the unit by pressing the I/O button. Disconnect the unit from the power
source when not in use.
3.4.
Cleaning and maintenance
a)
Pull out the mains plug before each cleaning and when the unit is not in use.
b)
Use only non-corrosive cleaning agents for cleaning the surfaces.
c)
After each cleaning, all the parts should be dried well before the unit is used
again.
d)
Store the unit in a dry and cool place protected from moisture and direct
sunlight.
e)
Do not spray the unit with a stream of water or immerse it in water.
f)
Make sure that no water enters through the ventilation openings in the casing.
g)
Clean the ventilation openings with a brush and compressed air.
h)
Perform regular inspections of the unit checking technical fitness and any
damages.
i)
Use a soft, damp cloth for cleaning.
j)
Do not use sharp and/or metal objects (e.g. a wire brush or metal spatula) for
cleaning, as these may damage the surface of the material from which the
unit is made.
k)
Do not clean the unit with acidic substances, medical products, diluent, fuel,
oil or other chemicals as this may cause damage to the unit.
l)
CAUTION! If there is concern that the equipment is contaminated with toxic,
pathogenic, radioactive, or infectious blood samples that are harmful to
health. Immediately sterilize the equipment according to current laboratory
methods.
m) Periodically lubricate the central part of the rotor.
TROUBLESHOOTING
Error code
E1
E2
E6
Meaning
Cover not closed
Motor drive does not work
Parameter setting error