• Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be
used in such environments:
Immunity
IEC 60601 test level
test
Conducted
3V rms
RF
At 0.15-80 MHz
IEC 61000-
4-6
6V rms
At ISM & Radio Amateur
Freq
Table continued on the following page.
Compliance
level
3V rms
At 0.15-80 MHz
6V rms
At ISM & Radio Amateur
Freq
39
Electromagnetic environment –
guidance
Portable and mobile RF communica-
tions equipment should be used
no closer to any part of the device,
including cables, than the recom-
mended separation distance calcu-
lated from the equation applicable to
the frequency of the transmitter.