Functioning
5.1.l - Alarms (Fig.11)
In the event of malfunctioning, the following alarms will
FIG. 11
be shown on the display :
Safety thermostat
When this alarm is activated, the "TS" alarm appears
on the display.
This indicates that the temperature in the chamber
exceeds 512°C; the oven heating system is turned off.
Reset the safety thermostat as described in the relevant
paragraph.
Chamber overheating
When this alarm is activated, the "TC1" alarm appears
on the display.
This indicates that the temperature in the chamber
exceeds 480°C; the oven heating system is turned off.
Turn off the oven and wait for the chamber to cool down.
Try again and call the technical support if the alarm
occurs again.
Chamber temperature probe open circuited or disconnected
When this alarm is activated, the "TC1" alarm appears
on the display.
This indicates that the chamber temperature probe is
open circuited or disconnected; the oven is turned off.
Call the technical support to check and/or replace the probe.
Chamber temperature probe short circuit
When this alarm is activated, the "TC1" alarm appears
on the display.
This indicates that the chamber temperature probe is
short circuited; the oven is turned off.
Call the technical support to replace the probe
EN - 24