Safety instructions
12K42
12K42
ErgoArm plus
ErgoArm plus
36 | Ottobock
12K44
12K44
ErgoArm Hybrid plus
ErgoArm Hybrid plus
The illustration shows that the plate and
brim of the lamination ring are level with
each other.
The electronics housing is fully inserted.
Modifying and servicing the forearm
Removing the elbow joint
1
Flex the arm fully. Turn the "finger wheel"
to minimum compensation and then extend
the arm fully.
Insert the fixation fork [1] to hold the cam
against the extension stop.
Lock the elbow joint in this position before
performing additional procedures.
Gently pop off both joint caps.
Carefully pry out the cable plugs.