7.6.3. Example: 3-point calibration for internal
80.00 °C
120.00 °C
160.00 °C
Menu level 1
Level 2
In the temperature range from 80 °C to 160 °C the calibration curve of the temperature
sensor (TT) is to be adjusted to the actual temperatures at measuring point (CT).
1. Set working temperature setpoint – SETPNT:
Refer to „Direct temperature setting " page 73
•
By pressing the key
example on the left: >SETPNT / 1 25.00°C<.
The integer digits flash
•
Change the value to 80.00 °C by pressing the keys
confirm by pressing the key
The decimal digits flash.
Confirm once more by pressing the key
•
The bath is heated up.
Wait for approx. 5 minutes until the temperature is constant.
2. Reading of temperature measuring device
Read the value of measuring point CT on the device and enter under menu item
>CALVAL X< by using the keypad.
>CALVAL 1< (79.70 °C)
>CALVAL 2< (119.50 °C)
>CALVAL 3< (159.30 °C)
3. Calibration
Press the
key if parameter is to be retained. Correction function for
parameters or values (prior to ENTER).
Setting is required only for the first calibration point.
Parameter level
Menu functions
control.
the circulator switches to the active >SETPOINT< see
(Example: <25>).
.
Ein ATC-Abgleich soll durchgeführt werden.
>NO< einstellen
•
Der Parameter blinkt, umschalten mit
+
•
Der Parameter blinkt, umschalten mit
+
Es wird >3. Punkt< -Abgleich durchgeführt.
91
and
and
.
.
.