5.

Setting of temperatures

5.1.
Using the pre-settings in the
Factory setting:
SETPNT 1
25 °C
SETPNT 2
37 °C
SETPNT 3
70 °C
Setting of temperatures
menu
Press the
key to call up the menu for temperature selection.
3 different working temperatures can be adjusted. Their values are freely selectable
within the operating temperature range.
The temperatures can be set in start or stop mode.
Press
key if a value is to be retained
Setting of working temperature in the
1.
Press the key
2.
Select SETPOINT 1 or 2 or 3 using the key
3.
Confirm by pressing the
The circulator uses the new working temperature value for temperature control.
Example: Adjustment/modification of the pre-setting of
"SETPOINT 3"
1. Press the
key.
2. Select SETPOINT 3 by pressing the
Example: SETPNT 3 / 70.00 °C
3. Keep the
(example: <70>)
4. Adjust value by pressing the
by pressing the
The decimal digits flash
Confirm once more by pressing the
Example on the left: SETPNT 3 / 85.00.
If the active setpoint (SETPNT) is changed, the new value is
immediately used for the control of the working temperature.
The heater control indicator flashes.
If the other two setpoints (not activated for control) are changed the
MENU has to be left by pressing the
digits have been confirmed
Notice: Refer to SETPOINT MAX / MIN in chapter
7.7.
MENU LIMITS
. The value flashes
key.
key pressed until the integer digits flash
key and the
key.
and can be adjusted if desired.
72
menu
or
.
key.
.
key to 85.00 °C and confirm
key.