Modification for using the measuring stand AVS or manual measurement inserts

Preparations
Filling
• Recommended maximum filling level with water as the bath liquid is
the top edge of the window.
• Recommended maximum filling level with oil as the bath liquid is
20 mm below the top edge of the window.
Draining
Press the mains switch to turn the transparent thermostat off.
Disconnect the power plug from the mains socket!
Slide a short piece of tube onto the drain port (17) and hold it into a
pail.
Unscrew the drain tap (18) and empty the unit completely.
Tighten the drain tap.
(Contents of delivery)
Bath lid B suitable for 2 measuring stands AVS
or
2 manual measurement inserts for capillary viscometers with
holders
CT 72/2, CT 72 TT
Lay centering insert for measuring stand AVS on the bottom of the
bath and push it in direction to edges
Lay the clamp on the rear panel and tighten both screws to fix the
centering insert to the bottom of the bath.
The insert for the bath lid is attached with six screws.
Each of the two openings can accept one measuring stand AVS
If necessary, immerse manual measurement inserts for visco-
meters in the opening.
Immerse viscometers with holders as per the instructions in the
corresponding operating manual for capillary viscometers.
Cover the unused openings with the lid.
CT 72/4:
Bath lid B suitable for 2 measuring stands AVS
and
2 manual measurement inserts for capillary viscometers with
holders