Preparations; Bath Liquids; Filling / Draining - Xylem SI Analytics CT 72/2 Manual De Instrucciones

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

3.

Preparations

3.1.

Bath liquids

Caution
3.2.

Filling / draining

Notice
Carefully read the safety data sheet of the bath liquid used, particularly with regard to
the fire point!
If a bath fluid with a fire point of ≤65 °C is used, only supervised operation is possible.
Water: The quality of water depends on local conditions.
Due to the high concentration of lime, hard water is not suitable for temperature
control because it leads to calcification in the bath.
Ferrous water can cause corrosion - even on stainless steel.
Chloric water can cause pitting corrosion.
Distilled and deionized water is unsuitable. Their special properties cause
corrosion in the bath, even in stainless steel.
SI Analytics GmbH recommends:
1. Water, in the mixing ratio: 70 % deionized water and 30 % tap water
Temperature range:
Important:
Using the Transparent Thermostat CT 72/2, CT 72/2 TT in the
Viscosity Automatic Sampler AVS Pro with water as the bath liquid,
the recommended temperature range is from 10 °C to 40 °C.
2. The use of silicone oil or mineral oil with a maximum viscosity of
30 mm
Temperature range:
3. Ethanol and Polyglycole
Temperature range:
Prevent humidity from entering into the immersion thermostat CT72.
Pay attention to the thermal expansion of bath oil during heating to avoid
overflowing of the liquid.
Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit, or before moving the unit.
Do not drain the bath fluid while it is hot!
Check the temperature of the bath fluid prior to draining (by switching the unit on
for a short moment, for example).
Always turn off the unit and disconnect the mains cable from the power source
before cleaning the unit, or before moving the unit.
Store and dispose the used bath fluid according to the laws for environmental
protection.
Preparations
5 °C bis 80 °C
2
/s,
-40 °C to 150 °C
-40 °C bis +10 C°
67