7.2.
MENU PROGRAM – creation, administration
Menu level 1
1 program
10 sections
D
RUN – the program has started or
PAUSE – the progress of the program has been interrupted by pressing the
key. While the time is stopped the temperature will constantly remain at
the last calculated setpoint
Continue with the
Termination / Interruption of a program
The program can be terminated any time by pressing the key
In case of power failure the program is interrupted.
The circulator switches to –OFF- .
If the AUTOSTART-function is activated the programmer starts again at the point
in time approx. 5 minutes prior to the interruption. However, an uncontrolled
change of the bath temperature has occurred.
The integrated programmer permits fast and easy programming of setpoint
temperature sequences. This temperature sequence is called program. A program is
composed of individual sections (STEP). The sections are defined by duration
(TSLICE) and target temperature. The target temperature is the setpoint (SETPNT),
which is achieved at the end of a section. The programmer calculates the temperature
ramp from the difference in time and temperature.
STEP
(Nr.)
SETPNT
(°C)
TIME
(hh.mm)
°C
200
150
100
75
50
25
10 20 30 40 50 60 10 20 30 40 50 60 10 [min]
Graph1
Sections without set value and time are skipped. They can be defined retroactively
and the integrated into the program.
Menu functions
key.
1
2
100
180
00:20
00.10
Step 1
2
4
80
4
6
180
75
00:20
00:50
6
8
.
8
75
00:20
t