Troubleshooting; Specifications - HoMedics BPA-200 Manual Del Usuario

Monitor automático de presión sanguínea
Ocultar thumbs Ver también para BPA-200:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

TROUBLESHOOTING

If any abnormality arises during use, please check the following points
POSSIBLE
SYMPTOMS
CAUSES
Unit does not turn on when
Batteries have
the START/STOP
run down.
button is pushed.
Battery polarities
have been positioned
incorrectly.
EE measurement error
Cuff has been placed
symbol shown on display or
incorrectly.
the blood pressure value is
displayed excessively low
Did you talk or move
(or high).
during measurement?
Shaking of the arm with the
cuff on.
Er error symbol shown on
Did you move during
display - measurement
measurement?
result exceeds
Shaking of the arm with
measurement range.
the cuff on.
E1 error symbol shown on
Air circuit abnormality. Cuff
display
tube may not be plugged
into monitor correctly.
E2 error symbol shown on
Inflation pressure exceeding
display
300 mmHg.
E3 error symbol shown on
Data error.
display
CA Appears on
Recalibration alert.
display
Note: If the unit still does not work, contact HoMedics Consumer
Relations. Under no circumstance should you disassemble or
attempt to repair the unit by yourself. Contact information for
HoMedics Consumer Relations Department can be found on
the warranty page.
29
CORRECTION
Replace them with
four new AA
alkaline batteries.
Re-insert the batteries in the
correct positions.
Wrap the cuff properly so
that it is positioned
correctly.
Measure again. Keep arm
steady during
measurement.
Measure again. Do not
move during measurement.
Check cuff connection.
Measure again.
Switch the unit off, then
measure again.
Remove and reload the
batteries.
Refer to page 25 for
more information.

SPECIFICATIONS

Four 1.5V DC (AA) batteries or
P P o o w w e e r r S S o o u u r r c c e e : :
supplied 12V DC AC adapter
M M e e a a s s u u r r e e m m e e n n t t
Oscillometric
M M e e t t h h o o d d : :
M M e e a a s s u u r r e e m m e e n n t t
Pressure: 40~250 millimeters mercury (mmHg)
R R a a n n g g e e : :
Pulse: 40~199 beats/minute
Pressure: ±3 mmHg
A A c c c c u u r r a a c c y y : :
Pulse: ±5% of reading
P P r r e e s s s s u u r r e e S S e e n n s s o o r r : :
Semi-conductor
I I n n f f l l a a t t i i o o n n : :
Pump driven
D D e e f f l l a a t t i i o o n n : :
Automatic Pressure release valve
M M e e m m o o r r y y C C a a p p a a c c i i t t y y : :
60 memories for each User (120 total)
A A u u t t o o - - s s h h u u t t - - o o f f f f : :
1 minute after last button operation
O O p p e e r r a a t t i i o o n n
Temperature: 50°F~104°F (10°C ~40°C)
E E n n v v i i r r o o n n m m e e n n t t : :
Humidity: 40~85% RH max.
S S t t o o r r a a g g e e
Temperature: 14°F~140°F (-10°C~60°C)
E E n n v v i i r r o o n n m m e e n n t t : :
Humidity: 10~90% RH max.
W W e e i i g g h h t t : :
0.80 lbs (without batteries)
A A r r m m C C i i r r c c u u m m f f e e r r e e n n c c e e : : 8.7"-18.1" (22-46cm)
4.5"(L) x 5.9"(W) x 3"(H)
D D i i m m e e n n s s i i o o n n s s : :
114mm(L) x 150mm(W) x 76mm(H)
Type BF
Device and cuff are designed to provide
special protection against electrical shocks.
Note: These specifications are subject to change without notice.
30

Capítulos

Tabla de contenido
loading

Tabla de contenido