ResMed AirCurve 10 Manual De Usuario página 19

Ocultar thumbs Ver también para AirCurve 10:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 50
Operating pressure range
ASV, ASVAuto:
CPAP
Supplemental Oxygen
Maximum flow:
Pneumatic flow path
Design life
Device, power supply unit:
Cleanable humidifier:
Air tubing:
General
The patient is an intended operator.
Humidifier performance
Mask Pressure
cm H
O (hPa)
2
3
4
10
20
25
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated
Air tubing
Air tubing
ClimateLineAir
ClimateLineAir Oxy
SlimLine
Standard
Heated air tubing temperature cut-out: ≤41°C
Notes:
• The manufacturer reserves the right to change these specifications without notice.
• The electrical connector end of the heated air tubing is only compatible with the air outlet at the device end and should
not be fitted to the mask.
• Do not use electrically conductive or antistatic air tubing.
• The temperature and relative humidity settings displayed are not measured values.
18
RH output % at 17°C
ambient
temperature
Setting 4
85
85
85
85
85
Material
Flexible plastic and electrical components
Flexible plastic and electrical components
Flexible plastic
Flexible plastic
4 to 25 cm H
4 to 20 cm H
15 L/min (CPAP, ASV); 4 L/min (ASVAuto)
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Humidifier
7. Device
8. Inlet filter
5 years
2.5 years
6 months
RH output % at 22°C
ambient
temperature
Setting 8
100
100
100
90
90
O (4 to 25 hPa)
2
O (4 to 20 hPa)
2
Nominal system output AH
Setting 4
6
6
6
6
6
Length
2 m
1.9 m
1.8 m
2 m
1
, BTPS
2
Setting 8
>10
>10
>10
>10
>10
Inner diameter
15 mm
19 mm
15 mm
19 mm
Tabla de contenido
loading

Tabla de contenido