ERROR CODES & TROUBLESHOOTING
Problem
Display shows an
abnormal result
Display is blank
when power is on
No response when
POWER is pressed
or when new
batteries have
been loaded
If an error code persists or is not remedied, please contact Customer
Care at 866-326-1313.
Possible Cause
Cuff position was not correct or
was not tight enough
Body posture was not correct
during measurement
Speaking, excessive movement,
angry, excited or nervous
anxiety during measurement
Batteries may be expired or
installed improperly
Incorrect operation; possible
strong electromagnetic
interference
Solution
Apply the cuff properly and attempt a
new measurement
Review the 'Taking a Measurement'
portion of this manual and attempt a
new measurement
Wait a period of time and attempt a new
measurement after calm has returned;
do not speak or move during
measurement
Check the polarity of the batteries and
reinstall if necessary; replace batteries
with new batteries
Remove batteries and wait a minimum
of five minutes; reinstall batteries and
attempt a new measurement