Technical Information
WIN# Read/
Type
Write
181
R/W
N
182
R/W
N
190
R/W
L
200
R
N
201
R
N
202
R
N
203
R
N
204
R
N
205
R
N
206
R
N
211
R
N
216
R
N
224
R
N
226
R
N
257
R
N
280
Meaning
Gauge Gas Type
(TPS-mobile equipped with 551
pump does not have this window)
Correction Factor for
*
Custom Gauge
Automatic Soft Start activation
(writable only if the pump is in STOP)
Pump current (mA dc); it reads the
actual value of the bus current supplied
by controller to electric motor
Pump voltage (Vdc); it reads the actual
value of the peak 3ph.voltage supplied
by controller to the electric motor
Pump power (W); it reads the actual
value of power supplied by controller to
electric motor
Driving frequency (Hz)
Pump temperature (°C)
Pump status
Error code (2^)
Controller heatsink temperature (°C)
Controller cooling air
temperature (°C)
Pressure reading
*
Format = X.X E XX
Rotation Frequency (rpm)
*
Gauge Status
Admitted
Values
0 = Not configured
1 = Nitrogen
2 = Argon
*
3 = Hydrogen
4 = Other
10¸999
Enable = 1
Disable = 0
0 = Stop
1 = Interlock
2 = Ramp
3 = Autotuning
4 = Braking
5 = Normal
6 = Fail
(bit description)
7 = Too High Load
6 = Short Circuit
5 = Over Voltage
4 = Override
3 = Run Up time
2 = Controller Over Temp
1 = Pump Over Temp
0 = No connection to pump
4 = RidUnknown
3 = GaugeOverRange
2 = GaugeUnderRange
1 = GaugeConnected
0 = NoGaugeConnected
Agilent TPS-mobile User Manual
Factory
Set
0
"1" at the first
START, then "0"