Table 8
Characteristic
Turbo Pump
TMP Flange
Forepump
Voltage
Base Pressure *
N
Pumping Speed (L/s)
2
He Pumping Speed (L/s)
H
Pumping Speed (L/s)
2
N
Max gas flow rate (sccm)**
2
He Max gas flow rate (sccm)**
H
Max gas flow rate (sccm)**
2
TMP Rotational Speed
Start-up time
Operating Position
Ambient Temperature
Bakeout Temperature
Protection Breaker
Installation category
Pollution Degree
Compliance with:
Indoor Use
Max altitude
Serial Communication kit
Storage Temperature
*
According to standard DIN 28 428, the base pressure is that measured in a leakfree test dome,
48 hours after the completion of test dome bake-out, with a Turbopump fitted with a ConFlat flange.
**
Note: values refer to air-cooled pump with:
Agilent TPS-mobile User Manual
TPS-mobile 230 V Technical Specification
measured in a leak-free test dome, 48 hours after the completion of
test dome bake-out, with a Turbopump fitted with a ConFlat flange
ambient temperature 25 °C or below
backing pump with pumping speed equal or above 5 m
X3581-64501, X3581-64503, X3581-64505, X3581-64507,
X3581-64509, X3581-64511, X3581-64513, X3581-64515,
X3581-64517, X3581-64519, X3581-64521, X3581-64523,
X3581-64525, X3581-64527, X3581-64529, X3581-64531,
X3581-64533, X3581-64535
305 FSQ
ISO 100
IDP-7, IDP-15, TS300, TS600, DS302, DS602
230 V / 50-60 Hz
< 1 x 10
-10
mbar (< 1 x 10
(according to standard DIN 28 428, the base pressure is that
and using the recommended pre-vacuum pump)
60600 rpm (1010 Hz driving frequency)
< 3 minutes (it can be longer in case of soft start active)
On a flat floor
+5 °C to +35 °C
max 75 °C on Inlet Flange
EN 55011 (Class A Group 1)
EN 61326-1
EN 61010-1
EN 1012-2
EN 12100
RoHS: EN 63000
-20°C to +70°C
-10
Torr)
250
255
220
375
450
450
14A
II
2
Only
2000 m
A-Plus
3
/h
251
Tabla de contenido
loading

Productos relacionados para Agilent TPS-mobile