Technical Information
Table 5
Characteristic
Turbo Pump
TMP Flange
Forepump
Voltage
Base Pressure *
N
Pumping Speed (L/s)
2
He Pumping Speed (L/s)
H
Pumping Speed (L/s)
2
TMP Rotational Speed
Start-up time
Operating Position
Ambient Temperature
Bakeout Temperature
Protection Breaker
Installation category
Pollution Degree
Compliance with:
Indoor Use
Max altitude
Serial Communication kit
Storage Temperature
*
According to standard DIN 28 428, the base pressure is that measured in a leakfree test dome,
48 hours after the completion of test dome bake-out, with a Turbopump fitted with a ConFlat flange.
246
TPS-mobile Technical Specification
9698433
TV551
CFF8
TS600
9698432
9698431
TV551
TV551
CFF8
CFF8
TS300
DS402
110 V / 60 Hz
1x10
-9
mbar (7.5x10
-10
Torr)
550
600
510
42000 rpm
<5 min.
On a flat floor
+5 °C to +35 °C
max 120 °C on Inlet Flange
25 A
II
2
EN 55011 (Class A Group 1)
EN 61326-1
EN 61010-1
EN 1012-2
EN 12100
RoHS: EN 63000
Only
2000 m
None
-20 °C to +70 °C
Agilent TPS-mobile User Manual
9698430
TV551
CFF8
DS302