Secondary Display
Most measurement functions have predefined range or measurement capabilities that can be displayed in the
secondary measurement mode. All math operations have predefined operations that are displayed on the secondary
display.
The table below shows the measurement functions which are available in the secondary display.
Primary Display
DCV
DCV
-
ACV
●
DCI
●
ACI
●
FREQ
-
– All specifications are ensured only under a single display.
– For ACI-ACV (primary or secondary) measurement, the ACV input signal is limited to 500,000 V×Hz.
– For DCI-ACV (primary or secondary) measurement, the ACV input signal is limited to 6,000,000 V×Hz.
– For DCV-ACV (primary or secondary) measurement, the DCV input signal is limited to 500 V when ACV input
signal is in 100 mV range. The ACV input signal must be greater than 50 mV.
– For ACI-DCV (primary or secondary) measurement operation frequencies, refer to the table below.
ACI-DCV
Slow/Medium
Fast
– For more information, please refer to
– For remote operation, refer to the DISPlay:WINDow2 commands in the Keysight EDU34450A Programming Guide.
Front Panel Operations
Press [Shift]> [Display] > 2nd Meas
Keysight EDU34450A User's Guide
Secondary Display
ACV
DCI
ACI
●
●
●
-
●
●
●
-
-
●
-
-
●
-
●
Measurement Operation Frequency
>500 Hz (600 Hz) / n × 50 Hz (60 Hz) for <500 Hz
>10 kHz / n × 1 kHz for <10 kHz
Measurement Speed
ON
| OFF.
FREQ
-
●
-
●
-
Consideration.
65