1.25
µS/cm
25.0
°C
C M=0.1000
A=0.0000
Save Calibration? Yes
16
After the calibration of the cell multiplier or slope calibration factor "M" i.e. cell constant
and the Adder or offset calibration factor "A" are displayed.
Select "Yes" to save the calibration values and the successful Calibration is confirmed
on the display.
The user gets the message "Re-install sensor" and "Press [ENTER]" on the display.
After pressing [ENTER] the M300 returns to the measuring mode.