9.3 Temperature Errors

Thermometry
Error Code
E 41
E 42
E 43
E 44
E 45
E 47
E 48
E 52
E 53
E 54
E 55
If a problem persists, discontinue use and return the device for servicing.
28
Description
Continuous temperature
over display limit (>43°C)
Continuous temperature
under display limit (<28°C)
No temperature probe
connected
Timeout on single
temperature
measurement
Continuous measurement
temperature circuitry not
operational
Single measurement
failed motion check
Single measurement
failed curve check
Single measurement over
display limit (>39°C)
Single measurement
under display limit
(<35°C)
Single measurement
start temperature out of
range (<10°C or >32°C)
Single measurement
temperature circuitry
not operational
Corrective Action
Re-calibrate
Re-calibrate
Attach temperature probe and repeat
measurement.
Allow probe to cool to room
temperature for 3-4 minutes, press
START, and then repeat measurement.
Internal malfunction. Return for
servicing.
Repeat measurement. Use
continuous mode if error is repeated.
As for E47 above
As for E47 above
As for E47 above
Allow probe to cool to room
temperature for 3-4 minutes, press
temperature START, and then repeat
measurement. Use continuous mode if
error is repeated.
Internal malfunction. Return for
servicing.