T T r r o o u u b b l l e e s s h h o o o o t t i i n n g g
P P r r o o b b l l e e m m
AST malfunction alarm
Kinked air hose alarm
Disconnected air hose
alarm
Remove from use alarm
EN
24
S S c c r r e e e e n n
C C a a u u s s e e
AST cable is not plugged in
AST cable is broken
AST circuit inside the
support surface is broken or
shorted
Bend or some obstruction in
the air hose that may cause
a compromise in the air flow
Air hose is disconnected
from the pump or there is a
leak in the system
Software or calibration data
corruption error
R R e e c c o o m m m m e e n n d d e e d d a a c c t t i i o o n n
1. Make sure that the AST
cable is connected.
2. Press the A A c c t t i i o o n n button
next to the A A l l a a r r m m o o f f f f
icon.
3. You may continue to use
the product. Turn off the
AST mode within three
seconds of canceling the
alarm.
4. If the condition persists,
contact a qualified
service technician.
Contact a qualified service
technician.
Contact a qualified service
technician.
1. Make sure that the air
hose is straight and air
flow is not obstructed.
2. Press the A A c c t t i i o o n n button
next to the A A l l a a r r m m o o f f f f
icon.
3. If the condition persists,
contact a qualified
service technician.
1. Make sure that the air
hoses are seated all the
way into the air hose
ports on the pump.
2. Press the A A c c t t i i o o n n button
next to the A A l l a a r r m m o o f f f f
icon.
3. If the condition persists,
contact a qualified
service technician.
Contact a qualified service
technician.
2941-009-005 Rev A.0