I I s s o o A A i i r r ® ® s s u u p p p p o o r r t t s s u u r r f f a a c c e e r r e e p p l l a a c c e e m m e e n n t t k k i i t t i i n n s s t t r r u u c c t t i i o o n n s s
If you ordered Model 2941 I I s s o o A A i i r r System, disregard these instructions and refer to the Model 2941 I I s s o o A A i i r r System
Operations Manual.
Use these instructions if you are replacing Model 2940 I I s s o o A A i i r r support surface with Model 2941-001-100 I I s s o o A A i i r r support
surface.
T T o o o o l l s s r r e e q q u u i i r r e e d d
•
None
P P r r o o c c e e d d u u r r e e
1. Loosen the tie down straps that secure the Model 2940 support surface to the bed frame.
2. Remove and discard the Model 2940 support surface.
3. Unpack the 2941-001-100 support surface and discard the packaging.
4. Setup the 2941-001-100 support surface follow the Model 2940 AO-SM-70-INT Operations Manual.
N N o o t t e e
• For instructions about how to use the Model 2941 I I s s o o A A i i r r Support Surface with the Model 2940 I I s s o o A A i i r r Pump, see the
AO-SM-70-INT Operations Manual. The content in this instruction sheet is a supplement to your Model 2940 I I s s o o A A i i r r
Operations Manual.
• Do not service the 2941-001-100 support surface with the Model 2940 Operations/Maintenance/Service Manuals.
• Do not launder the 2941-001-100 support surface.
• Use this instruction sheet for cleaning and disinfecting instructions of your 2941-001-100 support surface.
• To view your Operations or Maintenance Manual online, see https://techweb.stryker.com/.
For available service parts, contact Stryker Customer Service or Technical Support at: 1-800-327-0770.
S S u u m m m m a a r r y y o o f f s s a a f f e e t t y y p p r r e e c c a a u u t t i i o o n n s s
Always read and strictly follow the warnings and cautions listed on this page. Service only by qualified personnel.
W W A A R R N N I I N N G G
• Always secure the support surface to the bed frame to prevent the shifting of the product while in use.
• Always inspect for foreign objects between the support surface and the bed frame. Foreign objects may cause the
support surface to slide on the support platform.
• Do not use the support surface on a larger or smaller bed frame that does not fit the width, length, or thickness. This is to
avoid the risk of the support surface sliding, patient injury, or interference with moving parts of the bed.
• Do not use the support surface when gaps are present. The risk of entrapment can develop when the support surface is
placed on incompatible bed frames.
• Always check the CPR connector before use. Replace if visibly soiled. Direct skin contact with soiled material may
increase the risk of infection.
• Always inflate the product and unplug the pump from the mains before you clean or disinfect.
• Do not immerse the product in cleaning or disinfectant solutions.
• Always make sure that you wipe each product with clean water and dry each product after cleaning or disinfecting. Some
cleaning and disinfecting agents are corrosive in nature and may cause damage to the product if you use them
improperly. If you do not rinse and dry the product, a corrosive residue may be left on the surface of the product that
could cause premature degradation of critical components. Failure to follow these cleaning or disinfecting instructions
may void your warranty.
• Do not allow fluid to pool on the support surface. Fluids may cause premature degradation of components and may
cause unpredictable safety and performance of this product.
• Always disinfect the support surface between patients, to avoid the risk of cross-contamination and infection.
2941-009-020 Rev AA.2
1
EN