DISTAL ADAPTER
1.
Remove the outer PVA bag and roughly sand
the entire model to improve the bonding of the
remaining components. Be careful not to sand
through the lamination into the plaster model.
2. Place the model back in the transfer fixture.
3. Cover the center hole in the 4-Hole Adapter
(LLV-01041) with masking tape.
4. Attach the adapter to the socket using Fabtech
60 Second Adhesive.
5. Once the adhesive has set, remove the socket
from the transfer fixture.
ONE LINK
1.
Decide where to place the One Link.
•
WillowWood recommends medial placement.
Posterior placement is not recommended.
•
Must be placed distal to the One Sleeve Dummy
line.
•
The One Link must be able to sit flat against
the surface of the socket.
•
Mark the desired location on the socket.
Edges are smooth against the socket.
Example of an acceptable placement
7
of the One Link.