8.4 Measuring and setting ranges
8.4.1 Measuring parameter and working range for amperometric sensors
Measuring parameter
Cl
, ClO
2
2
O
3
8.4.2 Measuring parameter and working range for potentiostatic sensors
Measuring
Measuring
parameter
range
[ppm]
Cl
0.00 - 20.00
2
NH
0 - 100
3
ClO
0.00 - 1.00
2
HCl
0.0 - 30.0
O
0.00 - 1.00
3
Note
The measuring ranges depend on the set sensors and cannot be modified.
16
Measuring range
Accuracy
[ppm]
0.00 - 5.00
0.00 - 5.00
Resolution
Linearity
at 20 °C
[%] of full
[ppm]
scale
< 0.05
< 5
< 1
< 10
< 0.03
< 10
< 0.7
< 5
< 0.02
< 10
Temperature range
[%]
[°C]
± 10
+5 to +45
± 10
+5 to +45
Sensitivity drift
per 6 months
[%]
< 10
< 5
< 10
< 3
< 10
Product number
91835237
(314-011)
96687714
(314-013)
Temperature
Product
range
number
[°C]
96732268
-20 to +40
(314-021)
95700839
-20 to +40
(314-031)
95700837
-20 to +40
(314-041)
95700840
-20 to +40
(314-061)
95700838
-20 to +40
(314-071)