Actions During Operation; Display Options - Grundfos Conex DIA-G Instrucciones De Instalación Y Funcionamiento

Ocultar thumbs Ver también para Conex DIA-G:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 106
11.11.4 Damping the measured value display
The mean value generation of the display (and
therefore also of the current output) can be modified
in order to dampen a frequently changing measuring
signal. This stabilises the display of the measured
and the signal for the current output.
In the "Sensor 1"/"Sensor 2" submenu, the damping
can be entered in a range from 1-60 (factory setting:
1 = no damping).
Display S1/S2
– Press the [Up] or [Down] button to enter the
numerical value (= number of measured values
used for mean value generation).
– Save using [OK].

11.12 Actions during operation

11.12.1 Display options

Display with two parameters
Cl
Cl
2
2
0.30
ppm
0.35
ppm
Large display of measured values for both sensors.
The inverted header indicates the two sensors
aligned to the left and right.
The value for sensor 1, displayed on the left, is
shown in the first row, and the value for sensor 2,
displayed on the right, is shown in the second row
under the header.
Display with one parameter
Cl
2
0.30
ppm
Large display of the measured value. The selected
sensor is indicated in the header.
Alarm messages
A sensor-specific alarm refers to the sensor that is
flashing in the header. Alarms that do not relate to
sensors are displayed in the bottom row. There is no
flashing display for this type of alarm.
If several alarm messages are displayed, they can
be scrolled using the [Up] or [Down] button. They are
listed in chronological order. The selected alarm
message is displayed in inverted format. The system
then switches to the previously set display option.
Display for active alarms with two parameters
Cl
Cl
2
2
0.30
4.00
Alarm row 1
Alarm row 2
The sensors are displayed in the header aligned to
the left and right, with the measured values for the
sensors listed underneath. Depending on the
affected sensor, half of the header flashes in case of
an alarm.
Display for active alarms with one parameter
Cl
2
0.30
ppm
Alarm row 1
Alarm row 2
Large display of the measured value. The selected
sensor is indicated in the header.
Display with activated alarm and limit values
Cl
Cl
2
2
0.30
A
1 2
ppm
4.00
A
1 2
ppm
Activated alarm and limit values for sensor 1 or 2 are
displayed in the icons next to the relevant measured
values.
Sensors 1 and 2 each have two limit value relays,
and an alarm value can be activated for each sensor.
49
Tabla de contenido
loading

Tabla de contenido