Test Procedure - Stryker SERFAS Manual Del Usuario

Ocultar thumbs Ver también para SERFAS:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 151

Test Procedure

1.
Turn on the SERFAS Energy Generator.
2.
Turn on the electrosurgical analyzer.
Set the electrosurgical analyzer to 250 ohm load. (Refer to the electrosurgical
analyzer user guide for instructions on setting the output power load.)
3.
On the SERFAS Energy Generator, select the Cut level by using either
the Up/Down arrow buttons on the front panel, or the +/- pedals on the
footswitch.
4.
Activate the Cut mode at the selected level by pressing the yellow pedal
on the footswitch.
5.
Allow a few seconds for the measurement to stabilize.
6.
Take note of the displayed power in watts and the generator cut level.
The measurements should fall within the range specified in Table 1.
Table 1: SERFAS Generator Output Power at 250 Ohm Load
Cut
Mode
Level 1
Level 2
Level 3
Level 4
Level 5
Level 6
Level 7
Level 8
Level 9
Level 10
Level 11
Load
Output Power Range (Watts)
Minimum
250 ohms
250 ohms
250 ohms
250 ohms
250 ohms
250 ohms
250 ohms
250 ohms
250 ohms
250 ohms
250 ohms
Maximum
5
24
42
67
100
86
129
110
164
132
199
157
235
180
270
204
306
233
350
8
35
63
EN-19
Tabla de contenido

Solución de problemas

loading

Tabla de contenido