MP6025427-11A1 / A
needle stick injury.
Do not clean, resterilize, or try to extract the needle from the needle housing. An accidental needle stick may
occur.
Do not reuse sensors. Reuse of the sensor may cause damage to the sensor surface and lead to inaccurate
glucose values, site irritation, or infection.
Always make sure the sensor is firmly attached so that it does not come out of your body. Factors that may
increase the likelihood of the sensor falling off include increased physical activity (particularly for younger
patients) and improper taping technique.
Where to insert the sensor
CAUTION: Make sure the sensor site is at least 2.5 cm (1 inch) away from the insulin pump infusion
site or manual injection site. And when replacing the sensor, select a new site that is at least 2.5 cm
(1 inch) from the previous site.
CAUTION: Avoid the 5.0 cm (2 inch) around the navel to help ensure a comfortable insertion site, and
to help with adhesion.
Choose an insertion site that has an adequate amount of subcutaneous fat. The Enlite sensor has been
approved for use on the abdomen and buttock areas. See step 2 for a diagram of these areas.
Do not insert the sensor in muscle or areas constrained by clothing or accessories; areas with tough skin or
scar tissue; sites subjected to rigorous movement during exercise; or in sites on the belt or waistline to avoid
chafing or site irritation.
Storage and handling
Store the Serter in the released position, at room temperature, to maintain optimum product performance and
life.
Serter maintenance
The Serter is intended for single-patient or multiple-patient use.
WARNING: If blood comes in contact with any surface areas of the Serter used for multiple patients,
the Serter must be discarded. Dispose of the Serter according to the appropriate regulations for
contaminated devices.
WARNING: For multiple patients, always clean and then disinfect the Serter after each use.
Cleaning for single-patient and multiple-patient use
To clean the Serter, you will need the following materials: mild liquid soap (for example, Ivory
a soft-bristled toddler toothbrush, and a container.
To clean the Serter:
1. Prepare a mild liquid soap solution using 5 mL of mild liquid soap per 4 L of tap water.
2. Rinse the Serter under room temperature tap water for at least one minute. Continue rinsing until the
Serter is visibly clean. Make sure all hard-to-reach areas are rinsed completely.
3. While rinsing the underside of the Serter, press and release the mechanism on the underside of the Serter
to ensure that the underside of the Serter is completely rinsed with water.
4. Submerge the Serter in the mild liquid soap solution and soak it for at least 10 minutes.
5. Brush the entire surface of the Serter using a soft-bristled toddler toothbrush until visibly clean. Pay close
attention to brushing the hard-to-reach areas.
6. Rinse the Serter under running room temperature tap water for at least one minute, and until all visible
liquid soap is gone.
7. Shake off any excess water or moisture and place the Serter upright on a clean, dry, cloth to air-dry.
- 30 -
liquid soap),
®