EverFlo / EverFlo Q User Manual
•
Do not use this device near active high frequency surgical equipment and the Radio Frequency shielded
room of a Medical Electrical system for magnetic resonance imaging, where the intensity of electromagnetic
disturbances is high.
•
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into
service according to the EMC information provided in this manual. Contact your home care provider regarding
EMC installation information.
•
The EverFlo is designed to meet current electromagnetic compatibility requirements. However, if you suspect
the device's essential performance or significant functions are affected by or interfere with the normal
operation of other electronic devices (such as TV, radio or other household appliances), remove power and
try relocating the appliance or the device until the interference stops, or plugging the device into a different
power outlet controlled by a separate circuit breaker or fuse. If performance continues to be affected,
discontinue use and contact your home care provider.
•
Be aware that the electrical cord and/or tubing could present a tripping or strangulation hazard.
•
The No-Flow alarm is disabled when the low-flow meter is installed.
•
Do not modify this system or equipment in any way. Modifications could result in hazards to the user.
•
Place the device in a location so as to avoid pollutants and fumes.
•
The oxygen delivery settings of the oxygen concentrator should be periodically reassessed for the
effectiveness of the therapy.
•
If you feel discomfort or are experiencing a medical emergency while undergoing oxygen therapy, seek
medical assistance immediately to avoid harm.
•
To ensure that you receive the correct therapeutic amount of oxygen delivery according to your medical
condition, the Philips Respironics EverFlo device must be used:
-
Only after one or more settings have been individually determined or prescribed for you at your specific
activity levels
-
With the specific combination of parts and accessories that are in line with the specification of the oxygen
concentrator manufacturer and that were used while your settings were determined
•
Use only water-based lotions or salves that are oxygen compatible during setup or use during oxygen therapy.
To avoid the risk of fire and burns, never use petroleum or oil-based lotions or salves.
•
Do not lubricate fittings, connections, tubing, or other accessories of the oxygen concentrator to avoid the risk
of fire and burns.
•
Geriatric or any other patient unable to communicate discomfort, or hear or see the alarms while using this
device, may require additional monitoring.
•
Use only power cords supplied by Philips Respironics for this device. Use of power cords not supplied by
Philips Respironics may cause overheating or damage to the device and may result in increased emissions or
decreased immunity of the equipment or system.
•
Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
•
The use of incompatible parts or accessories can result in degraded performance.
8