Quick Start Programming: steps 1, 2 and 3
Notes:
P000 locks the unit. The unlocking value is 1954: all other values lock the unit.
Default settings in the tables below are shown with an asterisk.
Please refer to the full Instruction Manual to set up inputs, communications, and other
functionality.
1.
Configure all parameters to their factory settings via P999 Master Reset:
a.
Press
to activate PROGRAM mode.
b.
Press
twice to access parameter fields.
c.
Key in 999.
d.
Press
e.
Reset complete. (Note: Reset takes several seconds to complete.)
2.
Set the following parameters for a Quick Start:
3.
P001 Operation: Sets the type of
measurement required for the
application.
1
Material Level
2
Space
Values
3 * Distance (flange face to material)
P002: Material
1 * Liquids or slurries
Value (view only)
P003 Measurement Response: Determines
level change reaction.
1
Slow
2 *
3
Values
4
5
Fast
7ML19985QA85
to Clear All and initiate reset.
0.1 m/min
1 m/min
10 m/min
100 m/min
1000 m/min
SITRANS LR300 – QUICK START MANUAL
P004 Antenna: Identifies antenna
configuration.
240 * factory setting for all antenna types
Val-
241
rod + 50 mm PTFE extension
ues
242
rod + 100 mm PTFE extension
P005 Units: Specifies units used for
dimensional values.
1 * Meters
2
Centimeters
3
Millimeters
Values
4
Feet
5
Inches
P006 Empty: Distance in Units (P005) from
flange face to process empty point.
Range: 0.000 to 9999
Preset:10 m (or equivalent
Values
depending on units)
(Empty level can be set at any distance
desired, not just actual empty.
P007 Span: Sets the range levels to be
measured.
Range: 0.000 to 9999
Preset: 10 m (or equivalent
Values
depending on units)
(Full level can be set at any measurement
above the empty level.)
)
Page EN-7