Machine-Based Reprocessing Of The X-Smart Iq ® Contra-Angle (Thermal-Disinfection); Sterilization Of The X-Smart Iq ® Contra-Angle (Autoclaving) - Dentsply Sirona X-Smart IQ Instrucciones De Uso

Ocultar thumbs Ver también para X-Smart IQ:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 210
Cleaning, Disinfection and Sterilization
7.5.2
Machine-based Reprocessing of the X-Smart IQ
angle (Thermal-Disinfection)
Operation
Automatic cleaning /
disinfection
Cleaning at 65°C (149°F)
A
holding time 5 min.
Disinfection at 90°C (194°F)
holding time 5 min.
Ensure the X-Smart IQ
disinfection.
Remove any residual liquid with particle-free compressed air.
Drying is performed automatically during the drying cycle in the washer-disinfector.
Sterilization is required after Machine-based Reprocessing of the X-Smart IQ
angle.
7.5.3
Sterilization of the X-Smart IQ
For the calibration function refer to Endo IQ
Instructions.
Contra-angles must be checked for damage before they are reused. Damaged
products must be replaced.
If you only use X-Smart IQ
App), no calibration is required.
END-IFU-X-SMART-IQ-V16-WEB-EUR-Multilingual-2021-01
Use washer-disinfector conforming to EN ISO 15883,
e.g. Belimed WD, programme: neutral.
Use an authorized and suitable cleaning solution for
initial removal of organic and inorganic soils:
(recommended 0.5% (v/v) deconex 24 LIQ, Borer
Chemie or similar).
Use authorized and suitable neutralizing solution:
(recommended 0.25% (v/v) deconex 26 plus, Borer
Chemie or similar).
Deionized water (DI) or water that has this degree of
purity is recommended for cleaning.
®
Contra-angle attachment is completely dry after thermal
®
®
®
Handpiece in unconnected mode (without the Endo IQ
Remarks
Contra-angle (Autoclaving)
App User Manual -
6 Step by Step
®
Contra-
®
Contra-
®
31/52
Tabla de contenido

Solución de problemas

loading

Tabla de contenido