Cleaning And Maintenance; Additional Features; Hygiene Monitoring Systems - Terragene IC10FR Manual Del Usuario

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Additional Features

U s e
and
buttons to select the position number
corresponding to language code.
Finish printing language change
To save changes press and hold
seconds. A sound will be emitted indicating that process has
been completed successfully. A ticket with the new language
will be printed.

Cleaning and maintenance

Cleaning and decontamination of external surfaces
Disconnect the power supply cable and the USB cable from
the incubator. If the device is hot, wait until it has cooled down
before handling it.
Clean the external surfaces of the device using a microfiber
cloth moistened with a solution of mild dish washing detergent
and water. Wring the cloth so it is damp but not dripping
before cleaning, and wipe the outer surfaces of the incubator.
Afterwards, moisten a clean microfiber cloth with water
only, and repeat the procedure until all traces of detergent
are removed from the external surfaces of the device. After
cleaning, allow the Incubator to air dry for at least 1 hour before
connecting the power supply cable or the USB cable again.
This cleaning procedure can be followed whenever
considered appropriate.
This cleaning procedure must be followed every time a spill
occurs on any of the external surfaces of the incubator.
If further cleaning is required, or if you have doubts about
the cleaning agents you may use, please contact your local
distributor.
Maintenance
Bionova
Additional features

Hygiene monitoring systems

Chemdye
Chemdye
incubation position that allows the quantitative analysis
of the products destinated to evaluate cleaning and
contamination control.
IC10/20FR can run a quantitative analysis of PRO1MICRO
Hygiene monitoring system, adjusted by a BSA (Bovine
Seric Albumin) calibration curve. This analysis ensures the
traceability of the control of surface cleaning.
The method has a detection limit of 1 µg and a sensitivity of
0.3 μg. The result is indicated with a resolution of 0.1 μg, and
its accuracy is ± 0.5 μg (IC95%), in the range of 0 to 5 μg, and
±0.7 μg (IC95%), in the range of 5 to 10 μg with a range of 10 μg.
button for three
Reading Process:
1.Press and hold the button
reading program. A beep will sound. Then wait three seconds to
confirm the selection. The auto-reader will emit two beeps
indicating that a program has been selected.
2.Use the PRO1MICRO pen to take the sample from the surface
to be analyzed. For this, follow directions described in Figure 1
(Also explained on PRO1MICRO's directions for use).
3.Once the solutions contained within the device have been
correctly mixed, proceed with incubation (remember that the
reading cone must not contain the swab). For this, insert the
pen on the incubation/reading position (Figure 2). Make sure to
place PRO1MICRO as it is indicated on the auto-reader label.
When it is inserted in the correct position, it cannot be rotated
(Figure 2).
4.Press and hold the
the incubation and reading process (15 minutes 60 °C). The
lights
IC10/20FR Page 13
Do not clean the internal parts of the device.
Do not pour or immerse the device into any liquid. Do
not allow any liquid to run inside the device during
its cleaning.
IC10/20FR does not require routine maintenance.
®
PRO1MICRO reading system
®
PRO1MICRO reading system consists of an
®
Readings will be able to initiate once the incubation
temperature has been selected and is stable. The
lights corresponding to the selected temperature
(37 ºC or 60 ºC) will blink until the temperature
is reached, and then stay stable. PRO1MICRO
readings must be performed at 60 ºC.
for 1 second to enter the
button for 2 seconds to iniciate
and
will start blinking.
Tabla de contenido
loading

Este manual también es adecuado para:

Ic20fr

Tabla de contenido