A longer tube is supplied together with the stand in order to connect the bacterial filter
to the suction jar, so the suction jar can be hanged in front of the stand.

4.3 Secretion Jar Lid attachment

1)
Place Jar in vertical position
2) Insert lid with O-ring on Jar rim.
3) Clear Spring Clamp from outer rim, allowing lid to drop in place
4) Press Lid down firmly with handle.
5) Turn Spring Clamp into position.
6) Snap Spring Clamp down into holding position.
When using the VACUSON 40/60, the Secretion Jars must be in their holders at all
times! This is to avoid tipping and spillage which could destroy the pump!
Please follow the separate Instructions for the use of the disposable Monokit bottles (not included).

4.4 Suction Assembly Connection

1)
With Lid firmly secured, Hang Secretion Jar from support
mounting.
2)
Affix narrow Connector (1) onto Secretion Lid neck.
3)
Insert bacterial filter (3) into socket of main unit.
4)
Connect tube of baccterial filter (2) with bacterial filter and
push-on contact.
Insure that the VACUSON 40/60 is never operated without the Bacterial Filter,
otherwise the equipment could be destroyed!
5)
Hang Quiver (if included) on Stand holder.
6) Affix Overflow Connector to Secretion Jar Lid..
7) Affix Suction Tube to connector neck.
Do not forget to check the correct operation voltage on the rating plate!
8)
Plug in Control Unit Mains.
Refer to Function Control of the device's assembling, see 5.8 Function Control.
No. 31791 04/08
The System is Operational!