Safety instructions
3.7 Specific warnings
• Use a sterile single-use protective cover on the camera head and the coupler (option)
throughout the procedure. Check that the protective cover is not damaged.
• Check the compatibility of your endoscope with the manufacturer prior to use.
• This device is not sterile and is not sterilizable.
• This device was not designed to be used in an ionizing environment.
• When connecting the light cable or camera to the control channel, do not force the
connectors, as this may damage them.
• Ensure there is sufficient airflow to prevent any overheating within the device: at least 15 cm
clear all-round the device. Do not cover it and ensure that all of the device's feet are present
• If the power cord is damaged, immediately switch off the device. It is dangerous to operate
this device with a damaged cord.
• After the source has been used, when removing the fibre from its light guide, the
temperature of the metal part will be very high and could cause burns.
• The light intensity at the fibre outlet may cause eye damage. Be careful handling it when the
device is in use.
• To avoid any risk of electrical shock, this device must be connected only to a power system
equipped with protective grounding.
• Not suitable for use in the presence of a flammable anaesthetic mixture containing air,
oxygen or nitrous oxide.
3.8 Undesirable side effects
• There are no known undesirable side effect associated with the DUO Full HD (control unit
and camera head).
• Please refer to the user manuals for other devices used during surgical endoscopy or
diagnostic procedures (Gynaecology, Laparoscopy, Arthroscopy, ENT or spinal endoscopy).
3.9 Device moving
•The device may be moved using a trolley.
• Do not move the device while a procedure is under way.
3.10 Device assembly and disassembly
• The device must only be opened by a competent technician approved by the manufacturer.
• Only the fuses below the switch at the rear of the control unit may be replaced. The power
should be switched off and the fuses checked and replaced if necessary; only T 1.6 A-250 V UR
timedelay fuses should be used (only those marked UL/CSA).