CARE & MAINTENANCE
Proper care and maintenance of your blood pressure monitor will help ensure long life for the
product; improper use of the monitor may void the included warranty.
Always disconnect the cuff and tubing from the monitor prior to cleaning and storage.
Monitor Care and Maintenance
Do not use any liquids on the monitor; use a soft, dry cloth to clean the monitor as needed
Storage outside of stated storage temperature may result in measurement error or device
malfunction; storage environment temperature is: -4°F – 131°F (-20°C – 55°C);
humidity: ≤90% RH
Do not store in direct sunlight
Always keep the unit unplugged while not in use; remove the batteries if the unit will not
be used for an extended period of time
Do not attempt to disassemble your monitor; disassembling your monitor may void the
manufacturers warranty
Do not subject the monitor to strong shocks; take care not to drop the monitor
With proper care, the monitor will maintain performance for a minimum of 10,000
measurements or three years
No component of this device can be maintained by the user; the circuit diagrams,
component part lists, descriptions, calibration instruction or other information which will
assist the user's appropriately qualified technical personnel to repair those parts of
equipment which are designated repairable can be supplied
It is recommended that monitor performance be reviewed every two years or after service
or repair; please contact your service provider or call Customer Care