This chapter describes problems that you might encounter while using the Fluent Fluid Management System and how to solve them. For help with
errors not listed in this chapter, please follow the on-screen prompts.
During Setup
• If the In-FloPak is not detected, apply firm, even pressure on the In-FloPak to ensure it is seated properly and press 'Next'.
• If the Waste Bag is not detected, ensure tube connection is oriented vertically.
Loss of Suction
• Ensure the hysteroscope outflow tubing is connected to the hysteroscope.
• Ensure the buttocks drape tubing is connected to the buttocks drape.
• Ensure the outflow tubing is connected to the outflow port.
Visibility Issues
• Verify that the fluid bag clamp is open.
• To provide a tamponade effect, intrauterine pressure may need to be adjusted during the procedure.
• If performing a MyoSure procedure, advance the MyoSure Tissue Removal Device to the fundus and allow circulation of fluid through the blade
to clear field.
• Ensure the out-flow stop cock is fully open.
Poor Uterine Distension
• Avoid over dilation of the cervix. If the cervix is over dilated, use a second tenaculum to seal the cervix.
• Verify the fluid bag clamp is open.
• Ensure the pressure setting is adequate.
• Ensure the inflow tubing is not occluded or pinched.
Cutting Issues
• Ensure the MyoSure Tissue Removal Device is fully inserted into the Fluent Fluid Management System.