2.
Weighing mode
a)
Proceed in accordance to points 1: a) -d)
b)
Press the [U.WT] button to confirm the set value
scale range.
3.
Deleting information signal settings
a)
Proceed in accordance to points 1: a) -d) by entering
the number „0" sequentially while setting the upper
and lower limits or by pressing the [CLEAR] button.
3.3.8
PROGRAMMING
ACCORDING
PREFERENCES
1.
Automatic turning off
a)
Start the device and then press the „5" numeric
button four times to enter the programming mode.
„AOFF" will appear on the display [2].
b)
Press the [TARE] button to select the appropriate
minute value (2, 5, 8, 0).
c)
Press the [CLEAR] button to confirm the choice and
return to the normal operating mode, or
d)
Press the [TOTAL] button to confirm the choice and
move to the next settings.
2.
Zero-point tracking range
a)
In programming mode, press and hold the [TOTAL]
button. „TrAC" will appear on the display [2].
b)
Press the [TARE] button to select the appropriate
zero-point tracking range value (0, 1, 2, 3, 4). The
higher the value, the wider the range.
c)
Press the [CLEAR] button to confirm the choice and
return to the normal operating mode, or
d)
Press the [TOTAL] button to confirm the choice and
move to the next settings.
EN
3.
Zero display function
a)
In programming mode, press and hold the [TOTAL]
button. „ZEro" will appear on the display [2].
b)
Press the [TARE] button to select the appropriate
zero display range value (0, 1, 2, 3, 4). The higher the
value, the wider the range.
c)
Press the [CLEAR] button to confirm the choice and
return to the normal operating mode, or
d)
Press the [TOTAL] button to confirm the choice and
move to the next settings.
4.
Stability class range
a)
In programming mode, press and hold the [TOTAL]
button. „FiLt" will appear on the display [2].
b)
Press the [TARE] button to select the appropriate
stability range value (0, 1, 2, 3, 4, 5). The higher the
value, the shorter the stability display time.
c)
Press the [CLEAR] button to confirm the choice and
return to the normal operating mode, or
d)
Press the [TOTAL] button to confirm the choice and
move to the next settings.
5.
Zero resetting
a)
In programming mode, press and hold the [TOTAL]
button. „T.rAtE" will appear on the display [2].
b)
Press the [TARE] button to select the appropriate
zero resetting value (0, 1, 2, 3, 4, 5). The higher the
value, the more stable the zero point.
c)
Press the [CLEAR] button to confirm the choice and
return to the normal operating mode, or
d)
Press the [TOTAL] button to confirm the choice and
move to the next settings.
6.
Type of backlight
a)
In programming mode, press and hold the [TOTAL]
button. „bL" will appear on the display [2].
b)
Press the [TARE] button to select the appropriate
type of backlight (0 - auto, 1 - manual).
c)
Press the [CLEAR] button to confirm the choice and
return to the normal operating mode, or
d)
Press the [TOTAL] button to confirm the choice and
move to the next settings.
7.
Change of weight measuring units
a)
In programming mode, press and hold the [TOTAL]
button. „Unit" will appear on the display [2].
b)
Press the [TARE] button to select the appropriate
type of weighing units (0 - kg / g, 1 - lb).
c)
Press the [CLEAR] button to confirm the choice and
TO
USER
return to the normal operating mode, or
d)
Press the [TOTAL] button to confirm the choice and
move to the next settings.
8.
Case-lot weight conversion
a)
In programming mode, press and hold the [TOTAL]
button. „RU" will appear on the display [2].
b)
Press the [TARE] button to select the appropriate
conversion mode (0 - OFF, 1 - ON).
c)
Press the [CLEAR] button to confirm the choice and
return to the normal operating mode, or
d)
Press the [TOTAL] button to confirm the choice and
move to the next settings.
Caution: It is important to remember that to
preserve the correct settings and measurements of
the case-lot weight, gradually add more samples
to the samples placed on the scale. Otherwise, the
measurement result may be inaccurate.
Caution: The case-lot weight conversion function is
only effective after previous sampling.
9.
Checking the type of information signal
a)
In programming mode, press and hold the [TOTAL]
button. „PASS" will appear on the display [2].
b)
Press the [TARE] button to select the appropriate
information signal system (0 - internal, 1 - external).
•
Internal alarm - the alarm sounds when the
measured value is within the selected range.
•
External alarm - the alarm sounds when the
measured value is beyond the selected range.
c)
Press the [CLEAR] button to confirm the choice and
return to the normal operating mode, or
d)
Press the [TOTAL] button to confirm the choice and
move to the next settings.
Caution: After programming according to user
preferences, the device should be turned off and
then restarted in order to save the settings.
3.3.9 CALIBRATION
CAUTION: Calibration can only take place when the
weighing unit is a kilogram.
1.
Start the scale with the switch [11], make sure it is
zeroed.
2.
Press and hold the [ZERO] button, the display „5"
will show „CAL" and the display [2] will show the
weighing unit, which can be changed using the
numeric buttons.
3.
Set exactly the same weight on the scale [1] as shown
on the display [2].
4.
The display [2] will flash for a few seconds and then
return to the normal operation mode.
5.
Calibration is completed.
3.3.10 HANDLING THE BATTERY
1.
When the [G] symbol appears on the display [5], the
battery needs recharging.
2.
When the [G] symbol is displayed, the scale will turn
off after 10 hours.
12
3.
To fully charge the battery, it is recommended to
charge it for at least 12 hours.
4.
It is recommended to charge the battery every three
months if the scale is rarely used.
3.3.11 RS-232 TRANSMISSION
1.
Signal: MODE E1 A-RS 232C'S UART
2.
Format:
a)
Transmission speed: 2400 BPS
b)
Data bits: 8 bit
c)
Stop bit: 1 bit
d)
ASCII code
e)
Connector: 9-pin socket
f)
Output: Pin3
g)
Signal ground: Pin5
3.
Transmission in adding mode
a)
Press the [ADD] button
•
Record No. 1
Net 02000.0 g
U / W 000000 g
Pcs 000000
b)
Press the [ADD] button again
• Record No. 2
Net 03000.0 g
U / W 000000 g
Pcs 000000
c)
Press the [TOTAL] button
• Total
Net 05000.0 g
U / W 000000
3.3.12 ERROR CODES
ERROR CODE
E1, E2, E3
OL
Caution: If an error still appears after taking the above
steps, it is recommended to recalibrate the scale to solve
the problem.
3.4. CLEANING AND MAINTENANCE
a)
Use only non-corrosive cleaners to clean the surface.
b)
After cleaning the device, all parts should be dried
completely before using it again.
c)
Store the unit in a dry, cool place, free from moisture
and direct exposure to sunlight.
d)
Do not allow water to get inside the device through
vents in the housing of the device.
e)
The device must be regularly inspected to check its
technical efficiency and spot any damage.
Rev. 20.08.2019
Rev. 20.08.2019
f)
Do not leave the battery in the device if it will not be
used for a longer period of time.
g)
Use a soft, damp cloth for cleaning.
SAFE REMOVAL OF BATTERIES AND RECHARGEABLE
BATTERIES
The devices are fitted with 6V / 5.5Ah batteries. Recycle
batteries with the appropriate organisation or company.
DISPOSING OF USED DEVICES
Do not dispose of this device in municipal waste systems.
Hand it over to an electric and electrical device recycling
and collection point. Check the symbol on the product,
instruction manual and packaging. The plastics used
to construct the device can be recycled in accordance
with their markings. By choosing to recycle you are
making a significant contribution to the protection of our
environment. Contact local authorities for information on
your local recycling facility..
POSSIBLE REASON
SOLUTION
The weighing scale
Correct the
has not been set
scale settings.
correctly.
The scale was
Take the
turned on when
unnecessary
a load was on the
weight off
scale.
the scale and
restart the
scale.
If the [G] symbol is
Charge the
shown, the battery
battery.
is low.
Device overload
Take the load
off the scale.
13