Wireless technology
Technology used:
Connection types:
Frequency:
Max RF power output:
Operating range:
It is recommended that the machine is at a minimum distance of 0.43" (1.1cm) from the
body during operation. Not applicable to masks, air tubing or accessories.
IEC 60601-1 (Edition 3.1) classification
Class II (double insulation), Type BF, Ingress protection IP22.
Operating pressure range
AutoSet, AutoSet For Her, CPAP:
Maximum single fault steady pressure
Machine will shut down in the presence of a single fault if the steady state pressure
exceeds 30 cm H
2
than 1 second.
Pneumatic flow path
Pressure accuracy
Maximum static pressure variation at 10 cm H
ISO 80601-2-70:2015:
Tested with full face mask: ±0.5 cm H
26
O (30 hPa) for more than 6 seconds or 40 cm H
Bluetooth
SPP, iAP2, GATT
2402 to 2480 MHz
+4 dBm
10 m (Class 2)
4 to 20 cm H
1. Flow sensor
2. Blower
3. Pressure sensor
4. AAV (F20 connector only)
5. Vent
6. HumidX (N20, P10 connectors only)
7. Mask
8. Air tubing
9. Machine
10. Inlet filter
O (10 hPa) according to
2
0 (0.5 hPa)
2
O (4 to 20 hPa)
2
O (40 hPa) for more
2