Deleting Stored Results - Dräger CMS Instrucciones De Uso

Analizador con registro de datos integrado
Ocultar thumbs Ver también para CMS:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Operation
7.11

Deleting stored results

NOTICE
i
i
Reset date and time in datalogger after changing battery.
All stored results are deleted in the Setup menu.
ALL stored results are deleted in the process.
1. Move sliding switch from position 0 to
position 1. Then immediately press and
hold side button until the following display
appears:
Setup CMS
Read DataRec
2. Briefly press side button several times
until the following display appears:
Setup CMS
Delete DataRec
3. Press and hold side button until the
following display appears:
Delete DataRec ?
Yes/ No
The option to be set is underlined
Select Yes or No:
4. Briefly press side button.
Confirm selection:
5. Press and hold side button.
The
other
Setup
DataRecorder Read and Delete functions
can be selected by repeatedly pressing the
side button
Exit Setup mode:
6. Move sliding switch to position 0.
Dräger CMS Analyzer
options
and
the
Setup CMS
Read Data Rec
0
0
1
1
2
2
3
3
Setup CMS
Delete Data Rec
Delete Data Rec ?
Yes / No
Setup CMS
Delete Data Rec
01222837.eps
04422837_4en.eps
04522837_4en.eps
04422837_4en.eps
57
Tabla de contenido
loading

Tabla de contenido