SpEcificationS

Problem
Probable Cause
The SleepCube does
1. The Auto-ON feature has been disabled.
not start when
2. There is no power to the device.
breathing into the
mask.
3. Breath is not deep enough for the Auto-ON
feature to detect.
4. You are using a full face mask that has an anti-
asphyxia valve.
5. The air supply port plug is missing or not fully
inserted.
6. The humidifier chamber is not fully engaged
into the cradle or is missing.
The airflow has
1. The Auto-OFF feature has detected large
stopped unexpected
airflow because of loosely fitting mask.
during use.
2. During use your mouth opens and you begin
to mouth breathe.
The display shows a
An error has occurred in the device and requires
device fault.
servicing.
Symptoms of OSA
1. Air filter may be occluded.
have recurred.
2. Sleep apnea condition has changed.
Your skin becomes
1. Headgear is too tight or improperly adjusted.
irritated where the
2. Your mask may not be properly sized or is
mask touches your
not the most appropriate shape for you.
face.
Dryness of throat or
Inadequate humidity.
nose.
Condensation of
1. Your humidification setting is too high.
water is collecting in
2. The room temperature fluctuates from higher
the hose causing a
to lower levels through the night.
gurgling noise.
Air from flow
1. The air filter(s) is/are dirty.
generator seems to
2. The air inlet port is blocked.
be too warm.
3. The room temperature is too high.
4. The SleepCube is located near a heater.
Nasal, sinus or ear
You may have a reaction to air flow pressure.
pain, possibly with
runny nose.
sPeCifiCaTions
Size .........................................................................................................................................10.7 cm x 16.5 cm x 17.5 cm (4.2" H x 6.5" W x 6.9" D)
Weight ..................................................................................................................................................................................................... 1.22 kg (2.7 lbs.)
Electrical Requirements ..................................................................................................................................................................100-240V~, 50/60 Hz
Maximum Power Consumption ......................................................................................... 65 watts max from AC power source (flow generator only)
Pressure Range ............................................................................................................................................................................................. 3-20 cmH2O
Operating Temperature Range ............................................................................................................................................. 5°C to 40°C (41°F to 104°F)
Operating Humidity Range ............................................................................................................................................. 0% to 95% RH non-condensing
Operating Atmospheric Conditions ............................................................................................................................... Sea level to 9,000 feet (2743 m)
Storage & Transportation Temperature Range ............................................................................................................. -40°C to +70°C (-40°F to 158°F)
Storage & Transportation Humidity Range ..................................................................................................................... 0% to 95% RH non-condensing
Maximum Limited Pressure ............................................................................................................................................... 20 cmH2O under normal use
Remedy
1. Use the ON/OFF button to start and stop the device.
2. Verify that the power cord is firmly connected to the device and
the power outlet.
3. Breathe deeply in and out to start the SleepCube
4. Auto-ON may not work because your exhaled breath escapes
through the open valve. Use the ON/OFF button to start and stop
the device.
5. Make sure the air supply port plug is fully inserted into the unit.
6. Slide the humidifier chamber fully into the slot. Verify the latching
lever snaps into place.
1. Ensure you have a good mask fit; adjust the mask and headgear as
needed.
2. Contact your home care provider for a chin strap or different mask
which will help prevent mouth breathing.
If problem is not resolved with above remedies, disable Auto-OFF
feature in Advanced Menu options.
Contact your home care provider for service.
1. Clean or replace the air filter and relocate device away from
drapes and other dusty surfaces.
2. Contact your physician.
1. Loosen the headgear to reduce the contact pressure on your face.
2. Contact your physician or home care provider.
1. Add a DeVilbiss DV5 Heated Humidifier.
2. Increase the heater setting on the humidifier.
Reduce the humidifier heater setting and/or increase the temperature
of the room.
1. Refer to the Cleaning section for instructions on filter maintenance.
2. Unblock the air inlet.
3. Reduce the room temperature.
4. Locate the device away from the heat source so the air being
drawn into the device is at room temperature.
If the problem persists contact
your home care provider.
Discontinue use and contact your physician.