Problem
White, powdery
deposits appear
in the water
chamber.

Device messages

Code
MASK UNPLUGGED is
displayed on the screen.
HIGH FREQUENCY is
displayed on the screen.
is blinking in
the center of the screen.
The symbol
blinking in the center of
the humidification level
settings page.
US – 36
Possible cause
Traces of scale from
the tap water.
Description
The mask is
disconnected.
An overly high frequency
has been detected.
The device has detected
an operating error.
is
The device has detected
a humidifier operating
error.
Suggestion
Rub the traces with a sponge and liquid dish
detergent. Soak the water chamber in a solution of
1 part white vinegar to 9 parts water as described
under "Cleaning the water chamber", on page 33.
Rinse well with tap water. Dry the exterior with a
clean cloth.
Check
the
mask, the patient circuit, and the device.
This message disappears when any of the
four buttons on the front of the device is
pressed.
Press the device ON/OFF button twice.
Check the patient circuit connections. If
the problem recurs, contact your home
care provider.
Unplug the device from all sources of
electrical power. Plug it in again and turn
on.
If the error persists, contact your home
care provider.
If the device is connected to a battery,
the heated humidification feature cannot
be used and the symbol blinks.
Unplug the device from all sources of
electrical power. Plug it in again and turn
on.
If the error persists, contact your home
care provider.